3 Bars indicates Battery is full (100%)
2 Bars indicates 50% of the battery life is left
1 Bar indicates 25% of the battery life is left
Blinking battery casing indicates the need to replace
batteries with fresh ones as specified by manufacturer
Electrode is not in contact with solution or electrode
is failing
Replacement sensor is not connected properly to the
tester during sensor replacement
mV calibration error while attempting to calibrate tester to a
value which is out of range
Self-Diagnostic Messages
Low battery
indicator
Or / UrOr / Ur
Or / UrOr / Ur
Or / Ur
Over range /
Under range
signal
ErEr
ErEr
Er
.0.0
.0.0
.0
Error
Message
Electrode Replacement
You can replace the electrode module at the fraction of the cost of a new tester.
When the tester fails to calibrate or gives fluctuating readings in calibration standards,
you need to change the electrode.
1. With dry hands, grip the ribbed tester collar with electrode facing you. Twist
the collar counter clockwise (see picture A). Save the ribbed tester collar and
O-ring for later use.3. Align the four tabs on the new module so that they
match the four slots on the tester (see picture B).
2. Pull the old electrode module away from the tester.
3. Align the four tabs on the new module so that they match the four slots on the
tester (see picture B).
4. Gently push the module onto the slots to sit it in position. Push the smaller
O-ring fully onto the new electrode module. Push the collar over the module
and thread it into place by firmly twisting clockwise.
Note: It is necessary that you recalibrate your tester prior to measurement after an
electrode replacement.
SpecificationsSpecifications
SpecificationsSpecifications
Specifications
Large Screen TesterLarge Screen Tester
Large Screen TesterLarge Screen Tester
Large Screen Tester
ORPTestr10 / ORP Testr 10BNCORPTestr10 / ORP Testr 10BNC
ORPTestr10 / ORP Testr 10BNCORPTestr10 / ORP Testr 10BNC
ORPTestr10 / ORP Testr 10BNC
Range -999 mV to +1000 mV
Resolution 1 mV
Accuracy ±2 mV
Calibration Window Offset adjustment up to ± 150mV
Auto Off After 8.5 minutes from last key press
User Reset Yes
Non Volatile Memory Backup Yes
LCD Display Dual
Power Requirement 4 x 1.5V “A 76” micro Alkaline Batteries
Battery Life More than 500 hours
Operating Temperature 0 – 50 °C
Tester Dimension 6.5 “L x 1.5”dia. (165 x 38 mm)
Boxed: 220 x 60 x 50 mm
Weight 3.25 oz (90 gm)
Electrode Maintenance
1. Rinse the electrode with tap water or electrode storage solution after each
measurement.
2. In aggressive chemicals, dirty or viscous solutions, and solutions with heavy
metals or proteins, take readings quickly and rinse electrode immediately afterward.
3. If possible, keep a small piece of paper or sponge in the electrode cap – moistened
with clean water or electrode storage solution (NOT de-ionized water) – and close
the cap over the electrode.
Changing Batteries
1. Open battery compartment lid (with attached lanyard loop).
2. Remove old batteries; replace with fresh ones. Note polarity.
(shown in diagram below)
Applications
Water quality testing • pools • spas • cooling towers • water purification • chromate
reduction • cyanide oxidation • boilers • pulp bleaching • ozone systems • water and
wastewater treatment • and more!
Warranty
The waterproof ORPTestr10DJ and ORPTestr10BNC are warranted to be free from
manufacturing defects for 2 years and electrode module for 6 months. If repair, adjustment
or replacement is necessary and has not been the result of abuse or misuse within the
time period, please return the tester – freight prepaid – and correction will be made
without charge. Out of warranty products will be repaired on a charge basis.
Return of Items
Authorization must be obtained from your distributor before returning items for any
reason.
When applying for authorization, please include information regarding the reason the
item(s) are to be returned.
Note: We reserve the right to make improvements in design, construction and appearance
of products without notice. Prices are subject to change without notice.
68X068043 01/04 Rev 0
Rlbbed collar
Rotate collar
away from you
Insert Electrode Module
Electrode Module
Large
O-ring
Small O-ring
Small Tab Large Tab